Biological Cabinet With Class II B-2

NSW-205

Click on the icon to download details

Biological Cabinet With Class Ii B-2 Model (imported)

Specifications:

  • Accord with the regulation for Class II, Type B2 Biological Safety Cabinet in American NSF/ANS149.
  • The vertical lamina airflow has the patterns of 100% air exhaust to the outdoor or to the air exhaust system.
  • HEPA and ULPA with NO LEAKING (Patent No. ZL200520140347.2) in air supply and exhaust system, ensure the cleanliness of ISO (Class 100) or ISO Class 4 (Class 10).
  • The view screen of gradient of 10° make it comfortable to operate.
  • Non-obstructive air return mode in the front suction channel (Patent No. ZL200520140353.8).
  • V notch and humanized design of the shelf can make sure the protection of negative pressure and alleviate working tiredness.
  • The air supply and exhaust pipelines adopt PDI Control Valve with BELIMO Actuator, which can effectively reduce air disturb from the outside.
  • Interlocked light and sterilizing.
  • Anti-splash safety spare socket and junction valve.
  • Intelligent auto-supply system ensures the change of air volume less than 10% when the resistance of filter increases 50% and enhance safety.
  • Leak detection in cabinet, no leakage at 500Pa.
  • Leak detection in HEPA/ULPA, leaking ratio is ≤0.01% (scanned), ≤0.005% (not scanned).

Work Dimension

MODEL SIZE
W       D      H
SIZE - BSC-1, B2   970 x 600 x 620
SIZE - BSC-2, B2 1270 x 600 x 620
SIZE - BSC-3, B2 1570 x 600 x 620